Aff Mark 2

Customer opinion : Be the first one to note the product
+ Add note
Customer review : Be the first one to comment on the product
+ Add review
Reference AS..
Functionalities - Drop sensorInfrared sensor (optional)
Functionalities - KeyboardThe keyboard can be locked during infusion (can be disactivated)
Functionalities - Drug libraryMedecine library (60 drugs available, addition and removal possible)
Functionalities - AFFAll the tubing of the range has an anti-free flow system to prevent any uncontrolled flow
Functionalities - MemoryRecord of the last 100 alarms stored Memorisation of parameters after switching off (can be disactivated)
Functionalities - PrimePossibility of priming the tubing with the pump or without
Functionalities - RetrobolusPressure removed after occlusion. Volume infused after removing occlusion less than 0.2 ml
Functionalities - KVOFlow rate to Keep Vein Open triggered once the infusion is completed
Functionalities - Primary/secondary2 infusions programmed sequentially
Functionalities - Flow rate/volumeContinuous infusion in Flow rate/Volume to be infused
Functionalities - Volume/timeContinuous infusion : Volume to be infused / time
Functionalities - TPNTotal parenteral nutrition
Alarms- Deported alarmRS 232 connected to deported system Adjustable volume of the audible alarm with 3 levels Adjustable power of status light indicator with 3 intensity levels
Alarms - Other alarmsInfusion completed, high flow rate, low flow rate
Alarms - System errorsAutomatic test to check the functions of the pump is performed internally; if it is not functioning correctly, a system error is triggered
Alarms - Battery flatWarning for at least 15 minutes, when the battery is flat
Alarms - Door openDetection of the door opening during infusion
Alarms- Tubing absentPresence of tubing sensor
Alarms - air bubblesDetection of air bubbles by ultrasound sensor. Adjustable size of bubbles detected: 50, 150, 500 and 1000 μL. Cumulative of 1 mL over 15 minutes
Alarms - Downstream occlusionPressure sensor above the pumping mechanism (blocked filter,closed clamp)
Alarms- Upstream occlusionPressure sensor underneath the pumping mechanism with adjustable pressure of 200, 500 or 1000 mmHg
Alerts - Prolonged pauseAlert when left in “pause” status for longer than 2 minutes
Alerts - Mains power supply defectAlert on loss of mains power supply
Alerts - Near endConfigurable “near end” warning time for between 5 and 30 minutes
Alerts - Battery lowAutonomous for a further 30 minutes after “Battery low” alert occur
Infusion - BolusMacro: 1-999 mL/h/1-99 ml Micro: 1-99.9 mL/h/1-99 mL Adjustable maximum values
Infusion - KVO1-5 mL/h (1 mL/h increments)
Infusion - VolumeMacro: 1-9999 mL (1 mL increments) Micro: 0.1-999.9 mL (0.1 mL increments) Adjustable maximum values
Infusion - flow rateMacro: 1-999 mL/h (1 mL/h increments) Micro: 1-99.9 mL/h (0.1 mL/h increments) Accuracy of +/- 5% with tubing of the PFDP range Adjustable maximum values
Mechanical screenTransfl ective LCD display 128x128 pixels (63x63 mm), backlight with 5 intensity levels
Mechanical - power supply100-240 V AC, 50-60 Hz, 2 A max
Mechanical - IPX1Protection against drops of water falling vertically
Mechanical - Weight3 kg
Mechanical - Dimensions120x195x240 mm
Mechanical - CE Mark0459
SUPPLYNiMH technology; 7.2 V 1800 mAh 6 h autonomy at 25 mL/h Recharged in 4 h